8.6 Flag Priority List
The list defines the display priority in ascending order
- The Flags of the Treatment Group are displayed in parallel with the result flag
- The Flags of the Derived Group are represented separately in the detail and printout
No. Flag-Name NG-Ref Check name
5 High Abs >Abs Check: High Absorbance (Excessive Absorbance)
68 No Fluid Samp.B ISE Controller Flag: NO FLUID
90 No Isecal ISE.S ISE Controller Flag: NO STD
19 Air Fluid ISE.E
ISE Controller Flag: SEG FLUID
ISE Controller Flag: AIR FLUID
91 Air Isecal --
ISE Controller Flag: SEG STD
ISE Controller Flag: AIR STD
18 Ise Unstab ISE.N ISE Controller Flag: ISE UNSTAB
12 Reag Rng S1A.E Check: Reagent Range
9 Ep Unstab >React Check: Endpoint Unstable
56 High Act >Kin Check: High Activity / Low Activity
14 Curv Dir Std.E Check: Curve Direction
11 Non Linear >Lin
Check: Kinetic Readings
Check: Non Linear
13 >Repl Dev Dup.E Check: Duplicate Standard
16 Calc Error Cal.E Check: Calculation Error
80 Non Mono Mono.E Check: Curve Non Monotonic
20 Out of Rng Slop.E Check: Slope Out of Range (ISE)
84 Out of Rng Sig.E Check: Curve Out Of Range
17 >Std Dev SD.E Check: Standard Curve Deviation
22 Sol1 F Dev IStd.E Relativer Check des Solution-1 Faktors
6 Ag Excess >Proz Check: Antigen Excess
26 >Test Rng >Test
Check: Test Range (Teil 1)
Check: Test Range (Teil 2)
27 <Test Rng <Test
Check: Test Range (Teil 1)
Check: Test Range (Teil 2)
40 >RR >Ref Check: Reference Range
41 <RR <Ref Check: Reference Range
43 Cal Error Cal.E Check: Calculation Error (Sample Flag)
46 Calc Error Samp.? Result calculation out of range (Not used)
29 R 2(2s) S2-2Sa QC Check: R2 (2s)
35 R 1(3s) Q3SD QC Check: R1(3s)
36 R 1(2.5s) Q2.5SD QC Check: R1 (2.5s)
110 ? Cal -- Derived-Flag:Flagged Calibration
111 ? QC -- Derived-Flag:Flagged Quality Control
112 Cal Due -- Derived-Flag:Flagged Calibration Due
120 > > Treatment-Flag:Repeat
121 V v Treatment-Flag:Rerun Diluted
122 ^ ^ Treatment-Flag:Rerun Concentrated (Not used)