7.2.2.2 Message Header Record (H)
This record must always be the first record in a transmission. This record contains
information about the sender and receiver, instruments, and computer system whose
records are being exchanged. It also identifies the delimiter characters. The minimum
information that must be sent in a Header record is: H|\^&{RT}
The H corresponds to the record type, H=Header. The | (vertical bar) is used as the field
delimiter. The \ (backslash) is the repeat delimiter. The ^ (caret) is the component delimiter.
The & (ampersand) is the Escape delimiter. {RT} is the record delimiter i.e. <CR>.
The entire header record consists of the following fields:
H|\^&|||c111^Roche^c111^0.5.4.0509^1^1005|||||host|RSUPL^BATCH|P|1|200
51021152259{RT}
Field Name Reference Format
Comments (Data Content) H|7.1.
Record Type ID (H) 07.1.01 ST ‘H’
Delimiter Definition 07.1.02 ST The first character is the field delimiter, the second is the repeat delimiter,
the third is the component, and the fourth is the escape character i.e.
|\^&
Message Control ID 07.1.03 Not used / ignored by cobas c 111 instrument
Access Password 07.1.04 Not used / ignored by cobas c 111 instrument .
Sender Name or ID:
Name
^Manufacturer
^Instrument Type
^SW Version
^Protocol Vers.
^Serial Number
07.1.05 CM Name of the machine transmitting this message.
cobas c 111 instrument receives data: These fields will be logged but not
used to confirm the proper sender.
cobas c 111 instrument transmits data:
<Name > The analyzer name given
1
.
^<Manufacturer> Roche”
^<Instrument Type> “c111
^<SW Version > Instrument e.g. 1.2.3.1234
^<Protocol Vers.> Comm. Protocol 1
^<Serial number> Instrument e.g. 1235
Others 07.1.06
07.1.09
Not used / ignored by cobas c 111 instrument
Receiver ID:
Name
07.1.10 ST Name of the machine receiving this message.
cobas c 111 instrument transmits data: The host name given
2
.
cobas c 111 instrument receives data: This field will be logged but not
used to confirm the proper receiver (myself).
1
The name of the analyzer which has been setup in Utilities. See also Definitions – System.
Characters that can be used are printable ascii as defined in the appendix.
2
The name of the host computer which has been setup in Utilities. See also Definitions – System.
Characters that can be used are printable ascii as defined in the appendix.