cobas c 111
Host Interface Manual
Version
2
.1
52 / 93
7.2.2.9 Request Information Record
Record to request information from the other device. It is sent by the analyzer (within test
order request message) and the host (within result request message)
Q|1|^Sample1||ALL||||||||O
Note:
Up to SW v 2.0 the sequence number was missing. Corrected in SW v 2.0 and above
Field Name Reference Format
Comments (Data Content) Q|12.1.
Record Type ID 12.1.01 ST ‘Q’.
Sequence Number 12.1.02 NM Indicates the sequence numbers for the Test Request Record in the
current layer. Since this record is the layer that follows the Test Request
Record, it resets to ‘1’ each time a new Test Request Record is presented,
and then it is numbered consecutively; 1, 2, etc.
Starting Range ID
Number
12.1.03 CM Format: ^<Sample ID>
Element Max Length Format
Sample ID 23 ST
Element Description
Sample ID The sample ID is found on a barcode attached to the
test tube. It is read by a barcode reader in the analyzer. Total valid
character number is 23
It can be a manually entered number on the analyzer if barcode reading is
not enabled.
cobas c 111 instrument receives data:
The SampleID ‘ALL’ i.e. report on all known samples is not supported by
cobas c 111 instrument .
For ‘C’ requests (12.1.13) this field is ignored
Ending Range ID
Number
12.1.04 Not used / ignored by cobas c 111 instrument
Universal Test ID 12.1.05 NM cobas c 111 instrument receives data:
For ‘C’ = calibration requests the application code number (ACN)
For other requests the implementation assumes 'ALL' i.e. all available
results for the sample will be reported
Others 12.1.06
12.1.12
Not used / ignored cobas c 111 instrument
Request Information
Status Codes
12.1.13 ST Field Value Description
F The final patient sample result.
A Rejects the previous request and accepts the new
request. Information can only be requested through the Request Record
one at a time. To use this code, it is necessary to reject a current request
for information before requesting the next.
O Request for Test Request only (no results). Inquiry is
sent to a host from the analyzer using this code.
C Request a calibration result
EndOfRecord