9.15 Database Init Command - Host  Analyzer
Host Request EAC (Init database during next power cycle)
HOST 09:32:21,184 [ENQ]
c111 09:32:21,200 [ACK]
HOST 09:32:21,262 [STX]1H|\^&|||ASTM_SIM|||||c111|EAC^U07|P|1|200
50613010402[CR][ETX]44[CR][LF]
c111 09:32:21,293 [ACK]
HOST 09:32:21,309 [STX]2M|1|EQU|c111[CR][ETX]15[CR][LF]
c111 09:32:21,340 [ACK]
HOST 09:32:21,371 [STX]3M|1|ECD|12345|IN|N||C111-2509[CR][ETX]A8[
CR][LF]
c111 09:32:21,403 [ACK]
HOST 09:32:21,418 [STX]4L|1|N[CR][ETX]07[CR][LF]
c111 09:32:21,449 [ACK]
HOST 09:32:21,449 [EOT]
Analyzer Response EAR (the command was accepted or not)
c111 09:32:21,465 [ENQ]
HOST 09:32:21,465 [ACK]
c111 09:32:21,574 [STX]1H|\^&|||c111^Roche^c111^2.0.0.0710^1^3334
44|||||host|EAR^U08|P|1|20071210093221[CR][ETB]
B6[CR][LF]
HOST 09:32:21,590 [ACK]
c111 09:32:21,653 [STX]2M|1|EQU^RO^c111^1|c111|20071210093221|||[
CR][ETB]B9[CR][LF]
HOST 09:32:21,653 [ACK]
c111 09:32:21,715 [STX]3M|1|ECD^RO^c111^1|12345|IN|||C111-2509[CR
][ETB]50[CR][LF]
HOST 09:32:21,715 [ACK]
c111 09:32:21,778 [STX]4M|1|ECR^RO^c111^1|UN||C111-2509||[CR][ETB
]6C[CR][LF]
HOST 09:32:21,778 [ACK]
c111 09:32:21,809 [STX]5L|1|N[CR][ETX]08[CR][LF]
HOST 09:32:21,809 [ACK]
c111 09:32:21,840 [EOT]
OK indicates that the command was accepted, otherwise UN is reported (as in the example
above).