8.2 Calibration Flags
Nr. c111 Comment c111 Check
12
>Reag Rng
<Reag Rng
S1ABS abnormal
Description
During calibration the absorbance value of the lowest calibrator is below the lower defined
range or above the upper defined range.
13 >Repl Dev
DUPLICATE error
Description
Greater than replicate deviation.
Deviation between replicates exceed programmed limit. One or more replicates are
erroneous.
14 Curv Dir
STANDARD error
Description
Curve direction.
The direction of the calibration curve is incorrect / not as expected.
16 Calc Error
Calibration error (Calib Flag)
Description
Calculation error.
Slope or nonlinear standard curve cannot be calculated due to a calibration error.
17 >Std Dev
SD limit error
Description
Greater than standard deviation.
A point in the calibration fell outside the defined limits.
20 Out of Rng
Slope abnormal
Description
Slope out of range.
The slope for the specified electrolyte cartridge is outside of the acceptable/ programmed
range.
22 Sol1 F Dev
Internal Standard concentration abnormal
Description
The concentration of the Calibrator on the ISE unit was not within the expected range.
80 Non Mono
Monotony of curve
Description
Curve not monotonic.