.1
42 / 93
7.2.2.3 Message Termination Record (L)
This record occurs at the end of a message to indicate the end of a message.
L|1|N{RT}
Field Name Reference Format
Comments (Data Content) L|13.1.
Record Type ID (L) 13.1.01 ST ‘L’
Sequence Number 13.1.02 NM ‘1’
Termination Code 13.1.03 ST Input the value shown below.
N Normal termination
E Unspecified System error
EndOfRecord
7.2.2.4 Patient Information Record (P)
This record is used to transfer patient information to the analyzer (test order messages) or
to the host (result messages).
Note: On the cobas c 111 instrument there is no patient information stored or maintained
in the database, this record is therefore ignored when sent to a cobas c 111 instrument and
filled with a default patient information record when originating from the cobas c 111
instrument in e.g. transmission of result records.
In c111 SW V 2 the Laboratory ID field carries the manually entered sampleID part if in NPT
Mode (see chapter 2.5.3).
P|1||[SampleIDpart]
Field Name Reference Format
Comments (Data Content) P|8.1.
Record Type ID (P) 08.1.01 ST ‘P’
Sequence Number 08.1.02 NM Sequence number of the Patient Information Record in the message.
Always ‘1’.
Practice Patient ID 08.1.03 ST Not used / ignored by c111
Laboratory Pat. ID 08.1.04 ST Unique processing number assigned to the patient by the laboratory,
analyzer host
c111 receives data:
The field and record is ignored
c111 transmits data:
Only in NPT Mode the field carries the un-extended sampleID part (i.e. the
part entered by the user either manually or by barcode.
If not in NPT mode the field is left empty.
Rest of List 08.1.05
…
up to
…
08.1.35
Not used / ignored by cobas c 111 instrument
EndOfRecord