9. Appendix B - Communication Examples
In this section, some examples of communication on ASTM Upper Layer are shown.
This example text contains some non-realistic cases for standard clinical chemistry analysis
to show variable examples of communication.
In this section, the analyzer is shown as ‘c111’, External system is shown as ‘host’.
9.1 Order Query - Realtime / Analyzer to Host
c111 08:41:04,761 [ENQ]
HOST 08:41:04,761 [ACK]
c111 08:41:04,870 [STX]1H|\^&|||c111^Roche^c111^2.0.0.0710^1^3334
44|||||host|TSREQ^REAL|P|1|20071210084106[CR][E
TB]D6[CR][LF]
HOST 08:41:04,870 [ACK]
c111 08:41:04,917 [STX]2Q|1|^4456||ALL||||||||O[CR][ETB]01[CR][LF
]
HOST 08:41:04,917 [ACK]
c111 08:41:04,948 [STX]3L|1|N[CR][ETX]06[CR][LF]
HOST 08:41:04,948 [ACK]
c111 08:41:04,964 [EOT]
9.2 Order Query Response- Realtime / Host to Analyzer
HOST 08:41:06,979 [ENQ]
c111 08:41:06,995 [ACK]
HOST 08:41:07,073 [STX]1H|\^&|||ASTM-Host^V·6.61|||||c111|TSDWN^R
EPLY|P|1|20071210084105[CR][ETX]04[CR][LF]
c111 08:41:07,104 [ACK]
HOST 08:41:07,120 [STX]2P|1[CR][ETX]3F[CR][LF]
c111 08:41:07,135 [ACK]
HOST 08:41:07,198 [STX]3O|1|4456||^^^444^\^^^555|R||||||A||||||||
||||||O\Q[CR][ETX]6A[CR][LF]
c111 08:41:07,229 [ACK]
HOST 08:41:07,260 [STX]4C|1|L|Default·TS^^^^|G[CR][ETX]3F[CR][LF]
c111 08:41:07,291 [ACK]
HOST 08:41:07,307 [STX]5L|1|N[CR][ETX]08[CR][LF]
c111 08:41:07,323 [ACK]
HOST 08:41:07,323 [EOT]
9.3 Order Query Host Analyzer
HOST 08:38:07,040 [ENQ]
c111 08:38:07,055 [ACK]
HOST 08:38:07,134 [STX]1H|\^&|||ASTM_SIM|||||c111|RSREQ^REAL|P|1|
20050613010333[CR][ETX]73[CR][LF]
c111 08:38:07,165 [ACK]
HOST 08:38:07,212 [STX]2Q|1|^83720||ALL||||||||O[CR][ETX]1E[CR][L
F]
c111 08:38:07,227 [ACK]
HOST 08:38:07,243 [STX]3L|1|N[CR][ETX]06[CR][LF]