.1
48 / 93
7.2.2.7 Result Record
Record to transmit analytical data.
R|1|^^^111|1480.00|mmol/l||||F||UnitTest{RT}
Field Name Reference Format
Comments (Data Content) R|10.1.
Record Type ID 10.1.01 ST ‘R’
Sequence Number 10.1.02 NM Indicates sequence numbers for the Test Request Record in the current
layer. Since this record is the layer that follows the Test Request Record, it
resets to 1 for each occurrence of a new Test Request Record. It is
numbered consecutively; 1, 2, etc., for each occurrence of this record.
Universal Test ID:
^^^Test ID
^Treatment
type (and value)
^Pre-Treatment
type (and value)
^Result evaluation
type (and value)
10.1.03 CM Example: ^^^103^D
The first three elements for this format are empty. Currently Universal Test
IDs are not supported.
Element Max Length Format
<Test ID> Host Code Number
Ω
5 NM
^Treatment 3 ST
Element Description
Test ID, Profile ID (Host Code Number):
This is a key field for orders. When transmitting results, this field is left
empty, i.e. the relevant Test ID is part of the Result Record.
The analyzer identifies the test with 3-digit numbers.
Range: 1-999
NOTE: the Host Application Code is translated from and into the cobas c
111 instrument ACN space using a configurable table.
Treatment:
General: Specifies treatment to be done on instrument: (A)utodilution,
(D)ilution and factor, (C)oncentration and factor, etc., e.g. A, D100, etc.
(Roche specific field)
cobas c 111 instrument : Dilution or Concentration depending on the
treatment done according to the value entered in the application protocol
D / C / <empty>
<empty> - pipetting and testing is done using the standard analysis
parameters.
Pre-Treatment & Result Evaluation:
Not supported on cobas c 111 instrument
Ω
The record should be sent with this field left blank when the host has no Test Selections. See
example on page 80.