Field or control Description
Source
Select the I
2
C bus on which to trigger.
Trigger On Select the type of information on which to trigger.
Direction Sets the transfer direction on which to trigger (Read, Write, Either).
Available when Trigger On = Address or Address & Data.
Addressing Mode Sets the slave device address length (7 bits or 10 bits long).
Available when Trigger On = Address or Address & Data.
Address Sets the address pattern on which to trigger. The number of bits shown depends on the Address
Mode setting.
Tap the Binary or Hex field and use the A and B knobs to select and change the values. Or double-tap
on the field and use the virtual keypad to enter values.
Available when Trigger On = Address or Address & Data.
Data Bytes Sets the number of data bytes on which to trigger (one to five bytes). Use the A knob to change the
value.
Available when Trigger On = Data or Address & Data.
Data Sets the data pattern on which to trigger. The number of bits shown depends on the Data Bytes
setting.
Tap the Binary or Hex field and use the A and B knobs to select and change the values. Or double-tap
on the field and use the virtual keypad to enter values. See Binary, decimal, hex, and octal virtual
keypads on page 228.
Available when Trigger On = Data or Address & Data.
A, B knob controls Use the A knob to select (highlight) the digit(s) to change.
Use the B knob to change the value of the digit(s). Or double-tap on the field and use the virtual
keypad to enter values.
Mode & Holdoff
Trigger Mode determines how the instrument behaves in the absence or presence of a trigger event.
Holdoff sets the amount of time the oscilloscope waits after a trigger event before detecting and
triggering on the next trigger event.
For additional information on Trigger Mode, Holdoff, forcing a trigger, and the Trigger Frequency
Counter see Mode and Holdoff panel on page 226.
Act On Trigger
Set the actions the instrument takes when a trigger event occurs.
For additional information on the Act On Trigger panel see Act On Trigger configuration
menu on page 227.
LIN serial bus trigger settings panel
Use this menu (optional) to set up and display a LIN (Local Interconnect Network) serial bus waveform.
Note: Requires option SRAUT
O.
Menus and dialog boxes
3 Series Mixed Domain Oscilloscope Printable Help 207

Table of Contents

Other manuals for Tektronix MDO34

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO34 and is the answer not in the manual?

Tektronix MDO34 Specifications

General IconGeneral
BrandTektronix
ModelMDO34
CategoryTest Equipment
LanguageEnglish

Related product manuals