Field or control Description
Data Sets the data pattern on which to trigger. The number of bits shown depends on the Data W
ords
setting.
Tap the Binary, Hex, or ASCII field and use the A and B knobs to select and change the values. Or
double-tap on the field and use the virtual keypad to enter values. See Binary, decimal, hex, and octal
virtual keypads on page 228.
Available when Trigger On = Rx Data or Tx Data.
A, B knob controls Use the A knob to select (highlight) the digit(s) to change.
Use the B knob to change the value of the digit(s).
Mode & Holdoff
Trigger Mode determines how the instrument behaves in the absence or presence of a trigger event.
Holdoff sets the amount of time the oscilloscope waits after a trigger event before detecting and
triggering on the next trigger event.
For additional information on Trigger Mode, Holdoff, forcing a trigger, and the Trigger Frequency
Counter see Mode and Holdoff panel on page 226.
Act On Trigger
Set the actions the instrument takes when a trigger event occurs.
For additional information on the Act On Trigger panel see Act On Trigger configuration
menu on page 227.
SPI serial bus trigger settings panel
Use the SPI bus menu (optional) to set up and display an SPI (Serial Peripheral Interface) synchronous serial bus waveform.
Note: Requires option SREMBD.
Field or control Description
Source Select the SPI bus on which you want to trigger
.
Trigger On Select the type of information on which to trigger.
Data Bytes Sets the number of data bytes on which to trigger (one to sixteen bytes). Use the A knob to change the
value.
Available when Trigger On = MOSI, MISO, or MOSI & MISO.
MOSI Sets the data pattern on which to trigger. Double touch to open the editor to set the pattern.
Available when Trigger On = MOSI, or MOSI & MISO.
MISO Sets the data pattern on which to trigger. Double touch to open the editor to set the pattern.
Available when Trigger On = MISO, or MOSI & MISO.
A, B knob controls Use the A knob to select (highlight) the digit(s) to change.
Use the B knob to change the value of the digit(s).
Table continued…
Menus and dialog boxes
3 Series Mixed Domain Oscilloscope Printable Help 212

Table of Contents

Other manuals for Tektronix MDO34

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO34 and is the answer not in the manual?

Tektronix MDO34 Specifications

General IconGeneral
BrandTektronix
ModelMDO34
CategoryTest Equipment
LanguageEnglish

Related product manuals