Trigger Setup
Trigger Holdoff
Adjust holdof
f to obtain stable triggering when
the instrument is triggering on undesired
trigger events.
Trigger holdo
ff can help stabilize triggering,
since the oscilloscope does not recognize
new triggers during the holdoff time. When
the instrume
nt recognizes a trigger event, it
disables the trigger system until acquisition
is complete. In addition, the trigger system
remains dis
abled during the holdoff period
that follows each acquisition.
Holdoffs
Trigger Coupling
Trigger co
upling determines what part of the
signal is passed to the trigger circuit. Edge
and Sequence triggering can use all available
coupling
types: DC, AC, Low Frequency
Rejection, High Frequency Rejection, and
Noise Rejection. All other trigger types use
DC coupl
ing only.
Horizontal Position
When Delay Mode is on, use horizontal
position to acquire waveform detail in a
region
that is separated from the trigger
location by a signicant interval of time.
MDO4000 Series Oscilloscopes User Manual 81

Table of Contents

Other manuals for Tektronix MDO4054-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO4054-3 and is the answer not in the manual?

Tektronix MDO4054-3 Specifications

General IconGeneral
BrandTektronix
ModelMDO4054-3
CategoryTest Equipment
LanguageEnglish

Related product manuals