Acquire the Sign
al
How the Analog Acquisition Modes Work
Sample mode retains the rst sampled point
from each acqu
isition i nterval. Sample is the
default mode.
Peak Detect mode uses the highest and
lowest of all
the samples contained in two
consecutive acquisition intervals. This mode
only w orks with real-time, noninterpolated
sampling a n
d is useful for catching high
frequency glitches.
Hi Res mode calculates the average of all
the samples for each acquisition interval.
This mode a
lso only works with real-time,
noninterpolated sampling. Hi-Res provides
a higher-resolution, lower-bandwidth
waveform
.
Envelop
e mode nds the highest and lowest
record points over all acquisitions. Envelope
uses P eak Detect for each individual
acquisi
tion.
Average mode calculates the average value
for eac
h record point over a user-specied
number of acquisitions. Average uses
Sample mode for each individual acquisition.
Use ave
rage mode to reduce random noise.
Changing the A cquisition Mode, Record Length, and Delay Time
Use this procedure to change the acquisition mode.
1. Push Acquire.
MDO4000 Series Oscilloscopes User Manual 57

Table of Contents

Other manuals for Tektronix MDO4054-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO4054-3 and is the answer not in the manual?

Tektronix MDO4054-3 Specifications

General IconGeneral
BrandTektronix
ModelMDO4054-3
CategoryTest Equipment
LanguageEnglish

Related product manuals