PRI
Type:
Setting or query
Setting syntax:
PRI ON
PRI
OFF
Examples:
PRI ON
PRI OFF
Query syntax:
PRI?
Query response:
PRI ON;
PRI OFF;
Discussion:
This command enables the positive supply regulation in-
terrupt.
SRQ
is asserted when the positive supply changes
between any two of the three modes. These modes are con-
stant voltage, constant current and unregulated. The device
dependent serial poll status byte indicates the supply
causing the SRQ and the mode changed to. See Table 3-1.
The status message returned (with RQS asserted), as a
result of enabling the PRI interrupt, does not necessarily
show the instruments present status. Instead, it shows the
status which was latched at the time the interrupt occurred.
Use the command REG? to determine present status.
PRI
REGULATION?
Type:
Query
Query syntax:
REG?
REGULATION?
Query response:
REG
<number>, <number> ,<number>
Discussion:
This command provides a means of determining the reg-
ulation status of the three supplies. The three numbers
re-
turned apply to the negative, positive, and logic supplies in
that order. The numbers returned mean: (1) supply is in con-
stant voltage mode,
(2)
supply is in constant current mode,
(3)
supply is unregulated.
REGULATION?