TCPA300 and TCPA400 Performance Verification
5-16
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
TCPA300 Amplifier Test Record
Photocopy this form and use it to record the performance test results.
TCPA300 amplifier test record
Instrument Serial Number: Certificate Number:
Temperat ure: Relative Humidity %:
Date of Cal ibration: Technician:
Performance test Range, A/V Minimum Measured/calculated Maximum
Front-Panel Di splay Pass/Fail
DC Gain Accuracy (% Error) 1 -- 1 % +1%
5 -- 1 % +1%
10 -- 1 % +1%
50 -- 1 % +1%
50 (COMP) -- 1 % +1%
Bandwidth 1 .707 -- -- -- -- -- --
1 (100 MHz) .707 -- -- -- -- -- --
5 .707 -- -- -- -- -- --
10 .707 -- -- -- -- -- --
50 .707 -- -- -- -- -- --
50 (COMP) .707 -- -- -- -- -- --
AC Coupling
1
TEKPROBE Cable 1 1.05 Ap-p 1.5 Ap-p
BNC Coax Cable 1 1.05 Vp-p 1.5 Vp-p
Degauss Pass/Fail
Current Overload Pass/Fail
1
The AC Coupling test can be performed with either cable, but the units of measure for the results differ between the two.