Troubleshooting and Error Codes
3-24
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Correcting the Cause of an Error Code
After interpreting the error code, disconnect the probe from the circuit and
degauss the probe again. Take the measurement again. If the error code persists,
contact your Tektronix Service Center.
Shutdown Error
If all LEDs flash at the s ame time, a shutdown condition exists. If this occurs,
turn off the amplifier and disconnect the probe from the circuit under test.
Turning off the amplifier clears the error register.
A shutdown condition can be caused by exceeding the frequency/current
specifications. Verify that you are not exceeding the specifications and let the
probe and amplifier cool before using them again. In most cases, 15 minutes is
sufficient. If the measurement you were taking was within the probe and
amplifier specifications, degauss the probe and take the measurement again. If
the shutdown error persists, contact your Tektronix Service C enter.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TCPA300 SERIES and is the answer not in the manual?

Tektronix TCPA300 SERIES Specifications

General IconGeneral
BrandTektronix
ModelTCPA300 SERIES
CategoryAmplifier
LanguageEnglish

Related product manuals