EasyManuals Logo
Home>Tektronix>Amplifier>TCPA300 SERIES

Tektronix TCPA300 SERIES User Manual

Tektronix TCPA300 SERIES
205 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #84 background imageLoading...
Page #84 background image
Specifications
4-12
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
TCP404XL Maximum Measurement Times
Figures 4--8 through 4--10 show the maximum measurement time for the
TCP404XL probe, with varying duty cycles and temperatures. For more detail
on the relationship between measurement factors, refer to Measuring Noncontin-
uous Current with the TCP404XL Probe on page 3--6.
Measuring 750A at 50 _C ambient
Duty cycle (%)
0 5 10 15 20 25 30 35
0
2
4
6
8
10
12
Maximum measurement
time (min)
0A Continuous
100A Continuous
200A Continuous
300A Continuous
Figure 4- 8: Measuring 750A noncontinuous at 50 °C ambient temperature
Duty cycle (%)
0102030405060
0
5
10
15
20
25
30
Maximum measurement
time (min)
Measuring 600A at 50 _C ambient
70
0A Continuous
300A Continuous
400A Continuous
200A Continuous
Figure 4- 9: Measuring 600A noncontinuous at 50 °C ambient temperature

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TCPA300 SERIES and is the answer not in the manual?

Tektronix TCPA300 SERIES Specifications

General IconGeneral
BrandTektronix
ModelTCPA300 SERIES
CategoryAmplifier
LanguageEnglish

Related product manuals