EasyManuals Logo
Home>Tektronix>Amplifier>TCPA300 SERIES

Tektronix TCPA300 SERIES User Manual

Tektronix TCPA300 SERIES
205 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #96 background imageLoading...
Page #96 background image
TCPA300 and TCPA400 Performance Verification
5-6
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
DC Gain Accuracy
Use the following procedure to verify the DC gain accuracy of the amplifier.
Refer to Figure 5--2 and set up the test equipment as follows:
1. Connect the output of the amplifier to the DMM:
a. Connect a BNC Cable from the amplifier OUTPUT to the 50 Ω
precision feedthrough termination.
b. Attach the termination to a BNC-to-dual banana adapter.
c. Observe polarity and insert the dual banana adapter into the digital
multimeter DC voltage input.
2. Connect the probe input to the current source:
a. Connect the BNC end of the Calibration Adapter to a BNC-to-dual
banana adapter.
b. Observe polarity and insert the dual banana adapter into the current
source DC output. Do not connect the Calibration Adapter to the
amplifier at this time.
Current source
DMM
+--
BNC-to-dual
banana adapter
50 Ω Precision
termination
+--
Calibration
adapter
BNC-to-dual
banana adapter
Amplifier
50 Ω Coaxial cable
Figure 5- 2: Equipment setup for DC gain accuracy test
Test Equipment Setup

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TCPA300 SERIES and is the answer not in the manual?

Tektronix TCPA300 SERIES Specifications

General IconGeneral
BrandTektronix
ModelTCPA300 SERIES
CategoryAmplifier
LanguageEnglish

Related product manuals