TCPA300 and TCPA400 Performance Verification
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
5-17
TCPA400 Test Record
Photocopy this form and use it to record the performance test results.
TCPA400 amplifier test record
Instrument Serial Number: Certificate Number:
Temperat ure: Relative Humidity %:
Date of Cal ibration: Technician:
Performance test Minimum Measured/calcul ated Maximum
Front-Panel Di splay Pass/Fail
DC Gain Accuracy (% Error) -- 1 % +1%
Bandwidth .707 -- -- -- -- -- --
AC Coupling
1
TEKPROBE Cable 1050 Ap-p 1500 Ap-p
BNC Coax Cable 1.05 Vp-p 1.5 Vp-p
Degauss Pass/Fail
1
The AC Coupling test can be performed with either cable, but the units of measure for the results differ between the two.