TCP303 and TCP404XL Performance Verification
5-42
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
TCP303 Current Probe Test Record
Photocopy this form and use it to record the performance test results.
TCP303 test record
Instrument Serial Number: Certificate Number:
Temperat ure: Relative Humidity %:
Date of Cal ibration: Technician:
Performance test Range, A/V
Test
current, A
Minimum Measured/calculated Maximum
DC Gain Accuracy
1
(% Error) 5A/V 1 -- 3 % +3%
3 -- 3 % +3%
5 -- 3 % +3%
50 A/V 0.6 -- 3 % +3%
1.8 -- 3 % +3%
3.0 -- 3 % +3%
Rise Time 5A/V -- -- -- -- -- -- 23 ns
Bandwidth 5A/V -- -- -- .707 -- -- -- -- -- --
1
DC Gain Accuracy is tested at the 10 °Cto50°C specification. If you need to test the probe at 0 °Cto10°C, refer to the
specification limits on page 4- 1.