TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
5-1
Performance Verification Overview
The Performance Verification tests allow you to demonstrate that the Current
Probe Amplifier and Current P robes meet their specified levels of performance.
Since the TCPA300 and TCPA400 cannot operate without a current probe, a
calibration adapter is used to verify the amplifier(s) separately. The TCPA
Calibration Adapter is an optional accessory--see Replaceable Parts for part
numbers. After you have verified the amplifier performance independent of the
current probe(s), you can check the performance of current probes with
confidence.
Tolerances that are specified in these procedures apply to the amplifiers and
current probes and do not include test equipment error. The recommended
calibration interval is one year.
Performance Verification and Functional Checks
The warranted specifications for the amplifiers that are checked in the perfor-
mance verification procedures are listed in Table 5--1. See Table 4--1 in Specifi-
cations for warranted values. Functional checks are included for you to verify
proper amplifier operation.
Table 5- 1: Amplifier checks
Performance verification Functional checks
Amplifier
DC gain
accuracy
Bandwidth
Front-panel
display
AC coupling
Degauss
1
Overload
1
TCPA300
TCPA400
1
With probe attached
The warranted specifications for the probes that are checked in the performance
verification procedures are listed in Table 5--2. See Table 4--1 in Specifications
for warranted values.
Table 5- 2: Probe performance verification checks
Probe DC gain accuracy Bandwidth Risetime
TCP312
TCP305