Disassembly
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
7-9
8. Remove the screw that attaches the probe input connector to the front panel
amplifier board to the chassis. Pull the front panel assembly away from the
amplifier and set it aside. See Figure 7--6.
9. Remove the three screws that attach the amplifier board to the chassis.
10. Do in reverse steps 8 through 2 to install a new amplifier board.
Amplifier board
Front panel
Chassis
Screws
Torque screws to 4 in-lbs
when reassembling
Screws (3)
Torque screws to 8 in-lbs
when reassembling
Figure 7- 6: Removing the amplifier board