Self Test
4–6
TDS 340A, TDS 360 & TDS 380 Technical Reference
First, display the Cal menu.
Second, verify Pass
status for the
adjustment sections.
Third, run a signal path
compensation and verify
status is Pass.
Figure 4–2: Verifying adjustments and signal path compensation
8. Press Signal Path and then press the side menu button OK Compensate
Signal Paths. When compensation completes, the status message updates to
Pass or Fail in the main menu
9. Check that the word Pass appears under Signal Path in the main menu. (See
Figure 4–2.) If Pass does not appear, the oscilloscope has failed the
performance verification; return it to Tektronix for servicing.