XO FLEX Instructions for use Operation
26
Choose between three levels of spray water:
Increase amount of water with .
Decrease amount of water with .
To avoid necrosis, it is generally recommended to work with a “wet” spray that gives
at least 50 ml spray water per minute measured with the turbine hand-piece fitted!
In some cases e.g. when preparing a cavity that is not close to the pulp and you
want to minimize the spray aerosol it may be acceptable to use less water in the
spray.
Choose between three drive air levels:
Increase drive air with .
Decrease drive air with .
3.13 ULTRASONIC SCALERS
The ultrasonic scaler is used as described by the supplier.
For XO ODONTOSON scalers see below in section 3.14.
Lift the scaler forward and control it with the foot control.
Switch on the power (one step) with or .
Figure 24Foot control of scaler
See the power (in % of maximum) on the display.
Enable/disable irrigation water with X.
Choose between three levels of irrigation water:
Increase amount of water with .
Decrease amount of water with .
Choose between three power levels:
Increase maximum power with ▲.
Decrease maximum power with .
For details about XO ODONTOSON 360 scaler see below.