EasyManuals Logo
Home>Xo>Dental equipment>FLEX

Xo FLEX User Manual

Xo FLEX
95 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #33 background imageLoading...
Page #33 background image
XO FLEX Instructions for use Operation
33
Risk of contamination!
The disposable irrigation kit package has been sterilized by ethylene oxide gas. Sterility
cannot be relied upon if 1) the package has been opened or damaged or 2) the expiration
date stamped on the package has passed.
Appropriate sterile handling procedures must be observed to ensure and maintain sterile
conditions.
The disposable irrigation kits are for one-time use only. Do not re-sterilize!
XO Peristaltic Pump is delivered with two sample disposable irrigation kits. Additional kits can be
ordered from your XO distributor:
Pre-sterilized, disposable irrigation kit, 50 pcs (XO-055)
Pre-sterilized, disposable irrigation kit, 10 pcs (XO-056)
ATTACHING THE PUMP MODULE
Attach the pump module as shown in Figure 32.
Figure 32Attaching the pump module to the unit
After use, the pump module can be safely removed from the unit. It is not necessary to switch off
the unit before detaching the pump.
ATTACHING THE IRRIGATION TUBE
Release the pump head by turning the handle (1) counter-clockwise see Figure 33.
Figure 33Releasing the pump head
Attach the irrigation tubing to the pump head as shown in Figure 34.
Make sure the tube is held in place by the two plastic connectors.

Table of Contents

Other manuals for Xo FLEX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Xo FLEX and is the answer not in the manual?

Xo FLEX Specifications

General IconGeneral
BrandXo
ModelFLEX
CategoryDental equipment
LanguageEnglish

Related product manuals