XO FLEX Instructions for use Cleaning and infection control
63
To avoid accidentally exposing the patient to XO Suction Disinfection be careful not
to interchange XO Water Clean cartridges (white) and XO Suction Disinfection
cartridges (yellow)!
XO Suction Disinfection is not poisonous but is not intended for human intake.
Be careful not to spill the fluid as the painted surfaces may become stained.
Any spills must be wiped away immediately. Then clean the surface with a damp
cloth.
Do not use any other disinfectants for cleaning the suction system as this may
compromise the effectiveness of the system and damage vital parts in the suction
system!
Please note that failure to comply with this will affect XO’s warranty liability.
5.20 DISINFECTION OF UNIT WATER LINES
OVERNIGHT WATER TREATMENT
Normally, the overnight water treatment procedure should be used:
Step 1 (evening)
1. Fit the water disinfection instrument holder on the cuspidor bowl
2. Place all water-bearing unit instruments in the disinfection instrument holder on the
cuspidor bowl (make sure that the heating on the syringe is switched off)
3. Attach the cup filler outlet as shown in Figure 80
4. Activate the button marked (1) in Figure 80 and the water tank inside the unit is emptied
5. Wait for 3 minutes the instrument bridge display will show the remaining time of the
process
6. Switch the unit off and leave the unit instruments in the instrument holder on the cuspidor
Figure 80Activation of XO Water Clean (1) cuspidor (2) cup filler
Step 2 (morning):
1. Check that the unit instruments are still in the instrument holder on the cuspidor and that
the cup filler is attached
2. Switch on the unit
3. The water tank is filled and the unit instrument water lines are flushed with disinfection
liquid for 6 minutes the display will show the remaining time of the disinfection process