7.2 SERVICE MESSAGES AND REMEDIAL ACTIONS
Below you see an overview of the service messages that may appear on the instrument bridge
display.
Table 9 – Service messages
The unit will soon run out of XO
Suction Disinfection.
Make sure you have XO Suction
Disinfection cartridges (AN-354)
in stock.
The unit will soon run out of XO
Water Clean
Make sure you have XO Water
Clean cartridges (AO-980) in
stock.
Water lines are being flushed. MM
= remaining minutes to completion
of process.
Please wait – do not shut off
power or activate any buttons.
Water line flushing completed.
Unit operational again.
YYYY-MM-DD
Less than 30 days to next
preventive service and safety
inspection.
Call your XO authorized service
provider and require a preventive
service and safety inspection.
YYYY-MM-DD
Date for preventive service and
safety inspection is overdue.
Call your XO authorized service
provider and require a preventive
service and safety inspection.
Suction disinfection process
completed. Unit operational again.
The suction disinfection process is
ongoing. X = remaining time in %.
Please wait – do not shut off
power or activate any buttons.
WAIT MM
Water tank discharging water. MM
= remaining minutes to completion
of process.
Please wait – do not shut off
power or activate any buttons.
Pump detached or no connection
to unit.
Pump not connected to the
selected suspension/instrument
Attach pump correctly. Check
configuration of the pump.
If XO ODONTOCURE is activated
continuously the message “TOO
HOT” is displayed and the
instrument is turned off.
Depending on the temperature,
the light is ready for re-activation
after a few seconds.