XO FLEX Instructions for use Cleaning and infection control
61
Please note that manual cancellation of flushing is not possible.
In case of emergency you may switch off the unit, place the suction hoses in the holder and
switch on the unit again.
The suction nipple covers can be autoclaved.
If much blood has been sucked through the system, we recommend that the system be flushed
with at least 1 liter of cold water. After this, the system can be disinfected with XO Suction
Disinfection.
Disinfect the suction system with Dürr MD 555 1 2 times per week, if the unit is equipped with an
amalgam separator or if the above described procedure is not sufficient. Please see the
instructions provided by Dürr Dental.
Please note that XO Water Clean and XO Suction Disinfection cannot be activated simultaneously!
SUCTION FILTERS
Every evening the suction filters must be cleaned:
1. Disconnect the suction hoses from the unit see Figure 77
2. Eject the filters by pressing the button see Figure 78
3. Remove the filter cartridge from the filter holder – see Figure 79
4. Empty the filters
5. Wash the filters in a thermo-disinfector
6. For lubrication of O-rings, please use a silicon grease, approved for drinking water
applications and suitable for both EPDM and NBR rubber
Figure 77Removal of suction hoses and filters