ACF5000 FTIR ANALYZER SYSTEM | OI/ACF5000-EN REV. A 101
Manual start of automatic adjustment
NOTICE
The descriptions in this section refer to the adjustment of the analyzer
module FID and oxygen sensor.
Adjustment should only be carried out after the warm-up phase.
Automatic adjustment
Automatic adjustment can be carried out
as a zero point adjustment alone or
as an end point adjustment alone or
as a zero point and end point adjustment jointly.
Manual start of automatic adjustment
1 Select Automatic Adjustment menu:
MENU → Adjust → Automatic Adjustment
2 Zero adjustment alone: ZERO AUTOCAL
Span adjustment alone: SPAN AUTOCAL
Zero and span adjustment jointly: ZERO & SPAN AUTOCAL
Cancel automatic adjustment manually
The user can cancel the automatic adjustment during the procedure by
pressing the STOP softkey.
When automatic adjustment is stopped, the analyzer module is in an indefi-
nite state (as regards adjustment). For example, the zero point adjustment
may have been completed and calculated, but the end point adjustment has
not yet been carried out.
For this reason, automatic calibration will have to be restarted and allowed
to run to completion after any cancelation of automatic adjustment.