68 ACF5000 FTIR ANALYZER SYSTEM | OI/ACF5000-EN REV. A
Adjustment: Basics
NOTICE
The descriptions in this section refer to the adjustment of the analyzer
modules FID and oxygen sensor.
Controlling the adjustment
Controlling the adjustment
For the adjustment of the analyzer system, there are three methods of con-
trol:
manual adjustment,
automatic adjustment and
externally controlled adjustment.
Starting the adjustment
Manual adjustment is started manually via the analyzer system's display
and control unit.
Automatic adjustment is started at time intervals determined by the
internal clock or by an external control signal or manually via the analyzer
system's display and control unit.
Externally controlled adjustment is triggered by an external control sig-
nal.
Wait until the warm-up phase has ended
Adjustment should only be started after the warm-up phase.
Plausibility check during adjustment
If the analyzer system detects implausible values during adjustment (e.g. if
the end point value equals the zero point value), it cancels the adjustment
and issues an error message. The values stored for the last adjustment re-
main in effect.
Status signal
The "Maintenance mode" status signal is set during adjustment.