124 ACF5000 FTIR ANALYZER SYSTEM | OI/ACF5000-EN REV. A
Configuring status signals
Menu path
MENU Configure System Status Signals
Function
The configuration of the status signals is already defined when ordering the
analyzer system and set at the plant.
As a rule, it is not necessary to change this configuration during operation.
Selection
The following are available
Individual status signals, i.e. loss, maintenance request and maintenance
mode, as well as
Overall status signals.
NOTICE
If the configuration of the status signals is changed from overall status sig-
nals to individual status signal, the digital outputs of DO2 and DO3 of the
standard function block application status signals/externally controlled
adjustment occupied with limit signals are potentially overwritten with indi-
vidual status signals.
Further information on the status signals can be found in the "System sta-
tus: Status signals" section (see page 176).