ACF5000 FTIR ANALYZER SYSTEM | OI/ACF5000-EN REV. A 153
Changing analog output current range
Method
The current range of the individual analog outputs can be changed using the
parameter setting of the corresponding function blocks analog output. The
Technical Information "Function blocks – Descriptions and configuration"
contains complete information on the individual function blocks.
Menu path
MENU → Configure → Function blocks → Outputs →
Analog output
Changing current range
The current range is changed using the Output current range parameter.
Selection
The current range choices are 0 to 20 mA, 2 to 20 mA and 4 to 20 mA.
NOTICE
The output signal cannot be less than 0 mA and cannot be greater than 22 mA.
Current range limits
The output signal is limited to the range which is defined with the parameters
Lower Limit and Upper Limit. These parameters are factory-set to 0 mA
and 22 mA.