80 ACF5000 FTIR ANALYZER SYSTEM | OI/ACF5000-EN REV. A
Automatic Adjustment Check (AAC): Settings and export of the
QAL3 data
Function
The settings for the cyclical "Automatic Adjustment Check" are to be made
in the web page.
Menu path
ACF5000 Settings → QAL3
Password
A level 2 password is required to access this submenu.
Display
Measurements
The results of the AAC can be displayed in the Measurements submenu.
The data for the zero point or reference point can be displayed in a detailed
or simplified report. Selecting the required control chart to check the auto-
matic adjustment (CUSUM or Shewhart) opens a view of the relevant results
table.
If the results table for the QAL3 measurements has extensive results, use the
keys to switch between the individual pages.
The AAC results for a measuring component can be printed or exported
using the keys.
Export components
The AAC results for all measuring components are combined in a text file
that can be exported in the Export Components submenu.