122 ACF5000 FTIR ANALYZER SYSTEM | OI/ACF5000-EN REV. A
Delete system module
Sequence when removing system modules
When removing system modules from the analyzer system, the following
sequence must always be applied:
1 Delete system module in the software (instructions see below).
2 Remove system module from the analyzer system.
Delete an existing system module without replacement
1 Select Set system modules menu item.
The list of the system modules available in the system is displayed.
2 Select system module to be deleted (and not replaced).
3 Press REMOVE softkey.
This deletes irrevocably the parameter settings and the function block
configuration of this system module.
4 In the list, the system module now has the Deleted status.
5 Save configuration change with ENTER or discard with Back.