ACF5000 FTIR ANALYZER SYSTEM | OI/ACF5000-EN REV. A 125
Configuring Ethernet connection
Menu path
MENU Configure System Network TCP/IP Network
Function
The analyzer system can be integrated via the Ethernet 10/100/1000BASE-T
interface in an Ethernet network (with TCP/IP protocol).
Parameter
It depends on the DHCP settings what parameters need to be integrated:
DHCP on: network name (max. 20 characters, no empty and special charac-
ters),
DHCP off: IP address, IP address mask and IP gateway address.
Address
The IP address, IP address mask and IP gateway address need to be queried
from the system administrator.
NOTICE
Addresses of the TCP/IP categories D and E are not supported.
The address bits variable in the address mask may not all be set to 0 or 1
(broadcast addresses).
The IP address X9 may not be set to the group 10.0.0.X because this group is