ACF5000 FTIR ANALYZER SYSTEM | OI/ACF5000-EN REV. A 81
Edit components
Measuring components that must be checked during the AAC can be acti-
vated and configured in the Edit Components submenu.
A level 3 password is required to access this submenu.
Parameter Explanation
Active
QAL3 monitoring of measuring components can be activated and the fol-
lowing parameters can be configured.
sAMS Zero
Enter the value of standard devi-
ation from the zero point.
The unit corresponds to the unit of the
measuring range. System conditions
must be taken into account when using
the performance test data. Minimum
value = 3% of the measurement range.
sAMS Span
Enter the value of standard devi-
ation from the reference point.
Engineer
Enter name.
Comment
Enter comments