28 ACF5000 FTIR ANALYZER SYSTEM | OI/ACF5000-EN REV. A
Fuses
Function Rated value
–F80 Supply ground-fault circuit interrupter 30 A/30 mA
–F81 Fan or
air conditioner (option)
6 A or
16 A at 230 V, 20 A at 120 V
–F82 Lighting, service socket 6 A
–F83 Type 42 probe tube heater, PFE2 filter unit heater,
back-purging valves
6 A (only PFE2) or
10 A (PFE2 + probe 42, 230 V) or
16 A (PFE2 + probe 42, 120 V)
–F84 Heated sample gas line 16 A
–F85 ACF5000 electronics box AU1, ASP block heater, gas cell
heater
6 A
–F86 Air purifier, FTIR spectrometer, flow monitor, system
controller, power supply unit 24 V/5 A
6 A
–F87 Option "2nd measuring location": Type 42 probe tube
heater, PFE2 filter unit heater, back-purging valves
6 A (only PFE2) or
10 A (PFE2 + tube 42, 230 V) or
16 A (PFE2 + tube 42, 120 V)
–F88 Option "2nd measuring location": heated sample gas line 16 A
–F89 Option "2nd measuring location": heated selector valve
and heated sample gas line to analyzer cabinet
6 A
–F90 UPS supply ground-fault circuit interrupter 25 A/30 mA
–F91
to
–F99
Relay coils, contactor coils, solid-state relays, selector
solenoid valve (ceramic fuses)
T 2 A
High leakage current: 9 mA!