Aerosol Photometer
Aerosol Photometer - 2i 53
Setup
Figure 20: Setup Menu Screenshot
Aerosol Noise Suppression
i
Note
The Downstream Aerosol Noise Suppression (DANS) is
not available when the iProbe is connected to the
instrument. This prevents the user from accidentally
skewing results by suppressing actual readings in the
downstream mode.
Upstream Aerosol Noise Suppression (UANS)
The Upstream Aerosol Noise Suppression mode is
selectable by the operator when using the upstream
port during measurement, not while setting 100%.
When poor upstream mixing makes the measurement
of the upstream concentration difficult, the noise
suppression should dampen these variations.
An icon indicates when UANS is ON. The default
condition is ON.
Downstream Aerosol Noise Suppression (DANS)
The Downstream Aerosol Noise Suppression mode is
selectable by the operator when performing non-
scanning applications on each port. The DANS
performs applies an averaging function on the data
collected. It is not available when the probe is
connected.