Aerosol Photometer
Aerosol Photometer - 2i 70
APPENDIX D
Specifications
Main Unit Physical Characteristics
Dimension (L x
W x H)
15.2 cm (6 in) x 15.2 cm (6 in) x
33 cm (13 in)
Weight 8.6 kg (19 lb)
Operational Requirements
Power 115 VAC, 60 Hz, .5 Amps
220 VAC, 50 Hz, .25 Amps
Electrical Fuses Two Required, 5 x 20 mm, 2A,
250V, SLOW BLO
Operating Conditions
Ambient
Temperature
10ºC to 35ºC
Ambient
Humidity
5% to 95% (non-condensing or
icing)
Ambient altitude From 0 to 10,000 ft
Storage
requirements
-20 to 60º C
Less than 95% Relative Humidity
non-condensing
Aerosol Detection
Technique Near Forward Light Scattering
Photometer
Dynamic Range 0.0001 to 120 mg/m3
Sample Flow
Rate
Full flow through detection
28.3 liters per minute volumetric
Data Output Formats
USB Port 9600,8,1,N
Thermal Printer Thermal Paper, 57 mm (2 ¼ in)
wide