Aerosol Photometer
Aerosol Photometer - 2i 54
An icon indicates when DANS is ON. The default
condition is OFF.
i
Note
When the operator connects the probe, the DANS function
will automatically be disabled if it was previously selected.
This will preserve the integrity of the test results while
performing a filter scan.
Figure 21: Aerosol Noise Suppression Screenshot without Probe
Figure 22: Aerosol Noise Suppression Screenshot with Probe
Table 10: Noise Suppression Options List
Parameter Function Icon Limits
UANS
Dampens variations
in upstream aerosol
measurements
30 to 120 s
Increments
of 10 s
DANS
Applies an
averaging function
on the data
collected.
10 to 30 s
Increments
of 10 s
16 Feb 2012 09:59
Reagent
Al
a
rm
Set
P
o
in
t
PAO 0.001
Setup – Aerosol Noise Suppression
Alarms
Upstream (UANS) 30s
Downstream (DANS) 10s
Cancel Default
Save
16 Feb 2012 09:59
Reagent
Alarm Set Point
PAO 0.001
Setup – Aerosol Noise Suppression
Alarms
Upstream (UANS) 30s
Cancel Default
Save
PAO
Noise Suppression
UANS 30s
Default

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ATI Technologies 2i and is the answer not in the manual?

ATI Technologies 2i Specifications

General IconGeneral
BrandATI Technologies
Model2i
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals