Aerosol Photometer
Aerosol Photometer - 2i 55
Reporting Functions
Figure 23: Reporting Functions Screenshot
Table 11: Reporting Mode Options List
Parameter Function Icon Limits
Summary
Provides a summary
of the test to the
USB or Printer port.
N/A
Monitoring
Provides a summary
of the test to the
USB port with data
sets at user defined
intervals.
Increments
of 0.4 s
Continuous
Provides a %
leakage reading to
the USB port at
approximately a 1
second interval
N/A
16 Feb 2012 09:59
Reagent
Alarm Set Point
PAO 0.001
Setup – Reporting Functions
Alarms
Summary
Monitoring 0.8
Continuous
Cancel Default
Save
PAO
Reporting Functions
Summ.
Monit. 0.8
Cont.
Default