• Now you will need to electrically connect the M Series hood to the Base of the unit. To
complete this step all you will need to do is locate the electrical cord, extending from the
back of the hood, and plug it into the designated receptacle on the back of the base.
• The final step of the assembly process is to install the M Series power cord. The power
cord is installed by inserting the male “D” shaped plug into the corresponding female
receptacle located on the back of the device.
• Once you have installed the power cord, plug the unit into the appropriate wall outlet.
20.3 Special Notes
• The hood lamps have an “ON/OFF” switch so that they can be turned off if desired.
When the switch is in the ON position, the hood lamps will light when the base lamps
turn on.
Note: On combination UVA and UVB units, the hood will be equipped with a single
lamp type.
• With all combination units the UVA lamps will be the first four lamps in the unit. The
UVB lamps will be the last four. Therefore, if you are using the UVA lamps, position
your hands or feet at the front of the device. If using the UVB lamps, position your hands
or feet in the back half of the unit.
20.4 How to Position the Patient
The hands, feet or elbows should be placed through the arched exposure ports in the front shield
and laid directly on the acrylic guard that covers the lamps. Intensity is highest towards the
horizontal center of the unit. Rings or other jewelry may cause scratching of the shield and
should therefore be removed.
Note: With all combination units the UVA lamps will be the first four lamps in the unit. The
UVB lamps will be the last four. Therefore, if you are using the UVA lamps, position your hands
or feet at the front of the device. If using the UVB lamps, position your hands or feet in the back
half of the unit.
20.5 M Series Lamp Removal and Replacement
Lamps may need to be replaced due to burnout or because their energy output has decreased to
the point that you may no longer be receiving an effective treatment.
DO NOT REPLACE LAMPS INDIVIDUALLY. ALL LAMPS SHOULD BE REPLACED
AT THE SAME TIME TO AVOID UNEVEN TREATMENTS OR ERYTHEMA.
PLEASE CONTACT THE DAAVLIN SERVICE DEPARTMENT PRIOR TO
REPLACING YOUR LAMPS TO OBTAIN IMPORTANT INFORMATION ON HOW
THE OUTPUT OF YOUR UNIT WILL BE AFFECTED BY THE CHANGE.
Because of differences between brands, replace lamps with the same brand as originally
installed.