EasyManuals Logo
Home>Daavlin>Medical Equipment>NeoLux 3 Series

Daavlin NeoLux 3 Series User Manual

Default Icon
55 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #53 background imageLoading...
Page #53 background image
MNL-00008 [5] Proprietary and Confidential Page 53 of 55
Table 10 Recommended separation distances between portable and mobile RF communication equipment
and the ClearLink Controlled Phototherapy Equipment
The device is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the device can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the device as recommended below,
according to the maximum output power of the communications equipment
Rated maximum
output power of
transmitter (W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80
MHz
๐‘‘ = 1.2
โˆš
๐‘ƒ
80 MHz to 800 MHz
๐‘‘ = 1.2
โˆš
๐‘ƒ
800 MHz to 2.5 GHz
๐‘‘ = 2.3
โˆš
๐‘ƒ
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1: At 80MHz and 800MHz, the separation distance for the higher frequency range
applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
26.0 Warranty and Contact Information
26.1 Limited Warranty Policy
This Limited Warranty is provided to the original purchaser (the โ€œPurchaserโ€) of the Daavlin
device (the โ€œEquipmentโ€). Daavlin warrants Equipment to conform with the Equipment
specifications and be free from defects in material and workmanship during the device Warranty
Period. No warranty is made as to useful lamp life or as to reduction in ultraviolet output due to
any cause. DAAVLIN MAKES NO OTHER WARRANTIES OF ANY KIND WHATSOEVER
TO PURCHASER OR ANY THIRD PARTIES WITH RESPECT TO THE EQUIPMENT AND
HEREBY EXPRESSLY DISCLAIMS ALL WARRANTIES, EXPRESS OR IMPLIED,
INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE.
Product
Warranty Period
New Equipment
1 Year
Remanufactured Equipment
90 Days
Lamps
90 Days

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Daavlin NeoLux 3 Series and is the answer not in the manual?

Daavlin NeoLux 3 Series Specifications

General IconGeneral
BrandDaavlin
ModelNeoLux 3 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals