MNL-00008 [5] Proprietary and Confidential Page 37 of 55
Note: More information about the 1 Series Integrated Dosimetry Control System can be found
on page 17 under “Dosimeter Instructions: 1 Series”.
Your 1 Series has been specifically designed to deliver accurate and effective treatments from
two distinct treatment distances. Treatments can be administered directly on the acrylic surface
(0 in) or at a distance of 9 inches (22cm) away. Due to the inherent nature of phototherapy, the
light intensity will be much greater when treating directly on the acrylic surface (0 in) versus
treating from 9 inches (22cm) away. For this reason, it is necessary to select your desired
treatment distance before administering a treatment. By selecting your treatment distance, 0 in or
9 in, you are informing the sensor how it should be evaluating and calculating the light output.
After unlocking your Integrated Dosimetry Control System (See “Unlocking the Device” on
page 15) you will need to select a treatment distance of either “0 in” or “9 in”. Pressing the
“Power” button will allow you to toggle between these two options and pressing “Enter” will
lock in your desired treatment distance.
22.5.2 Calibration Output Certificate
Here at Daavlin we measure the power output of each and every unit that we manufacture before
allowing it to leave our facilities. Included with your literature you will also find a “Calibration
Output Certificate” that contains your device’s factory measured power output. Since your 1
Series has been designed to deliver treatments from two specific distances, 0 inches and 9 inches
(22cm), your 1 Series will have two outputs noted in the “Current Value” box of your certificate.
The higher of the two output values corresponds to the “0 in” setting and the lower of the two
values corresponds to the “9 in” setting.
Note: When calculating your treatment backup time, be sure to use the correct output value for
the distance that you wish to treat from. More information about calculating backup times can be
found on page 17 under “Dosimeter Instructions: 1 Series”.
22.6 Special Notes for 1 Series Units Equipped with the Digital Timer
22.6.1 Treatment Distances and Your Calibration Output Certificate
Your Daavlin 1 Series has been designed to administer treatments at two distinct treatment
distances. Treatments can be administered directly on the acrylic surface, making the treatment
of hands and feet comfortable and easy, or from a distance of 9 inches (22cm) away for
conveniently treating the face or other localized areas of the body.
Due to the inherent nature of phototherapy, the intensity of the ultraviolet light will be greater
when treating directly on the acrylic surface versus treating from 9 inches (22cm) away. In order
to insure accurate and effective treatments it is very important that you calculate your treatment
times using the power output that is appropriate to your desired treatment distance.
Here at Daavlin we measure the power output of each and every unit that we manufacture before
allowing it to leave our facilities. Included with your literature you will also find a “Calibration
Output Certificate” that contains your device’s factory measured power output. Since your 1
Series has been designed to deliver treatments from two specific distances, directly on the acrylic
and 9 inches (22cm) away, you will have two outputs noted in the “Current Value” box of your

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Daavlin NeoLux 3 Series and is the answer not in the manual?

Daavlin NeoLux 3 Series Specifications

General IconGeneral
BrandDaavlin
ModelNeoLux 3 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals