24.2.3 High-Level Disinfection
Follow a high-level disinfection protocol between the use of the device on different patients. Use
a high-level disinfectant, such as Steris Corporation’s Resert XL-HLD, and follow the
manufactures guidelines. See also “FDA-Cleared Sterilants and High Level Disinfectants with
General Claims for Processing Reusable Medical and Dental Devices” available at:
https://www.fda.gov/medical-devices/reprocessing-reusable-medical-devices-information-
manufacturers/fda-cleared-sterilants-and-high-level-disinfectants-general-claims-processing-
reusable-medical-and
24.3 Resetting Your Lamp Hours
Whenever the lamps are changed in your device, it is important to reset the lamp age to zero and
perform calibration.
If you need to reset lamp hours and/or perform calibration, please contact the Daavlin service
department at 1-800-322-8546 for step-by-step instructions.
25.0 EMC Guidelines
Table 7 Electromagnetic Emissions
Emissions Test
Compliance
EMC Environment Guide
RF Emission Following
CISPR 11 (EN 55011)
Not
Applicable
RF Emission Following
CISPR 11 (EN 55011)
Not
Applicable
Limits for Harmonic
Current Emissions
Following IEC 61000-
3-2
Class A
The device is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
Limitation of Voltage
Changes, Voltage
Fluctuations, and
Flicker Following IEC
61000-3-3
Complies
The device is suitable for healthcare environment operation
in hospitals and clinic
Table 8 Electromagnetic Immunity
Emissions Test
IEC 60601 Test Level
Actual Level
Electrostatic discharge
immunity test following IEC
61000-4-2
+/- 8kV (conductive
surfaces, coupling planes)
+/- 2kV, +/- 4kV, +/- 8kV,
and +/- 15kV (non-
conductive surfaces)
+/- 8kV (conductive
surfaces, coupling planes)
+/- 2kV, +/- 4kV, +/- 8kV,
and +/- 15kV (non-
conductive surfaces)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Daavlin NeoLux 3 Series and is the answer not in the manual?

Daavlin NeoLux 3 Series Specifications

General IconGeneral
BrandDaavlin
ModelNeoLux 3 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals