24.2.3 High-Level Disinfection
Follow a high-level disinfection protocol between the use of the device on different patients. Use
a high-level disinfectant, such as Steris Corporation’s Resert XL-HLD, and follow the
manufactures guidelines. See also “FDA-Cleared Sterilants and High Level Disinfectants with
General Claims for Processing Reusable Medical and Dental Devices” available at:
https://www.fda.gov/medical-devices/reprocessing-reusable-medical-devices-information-
manufacturers/fda-cleared-sterilants-and-high-level-disinfectants-general-claims-processing-
reusable-medical-and
24.3 Resetting Your Lamp Hours
Whenever the lamps are changed in your device, it is important to reset the lamp age to zero and
perform calibration.
If you need to reset lamp hours and/or perform calibration, please contact the Daavlin service
department at 1-800-322-8546 for step-by-step instructions.
25.0 EMC Guidelines
Table 7 Electromagnetic Emissions
The device is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
+/- 8kV (conductive
surfaces, coupling planes)
+/- 2kV, +/- 4kV, +/- 8kV,
and +/- 15kV (non-
conductive surfaces)
+/- 8kV (conductive
surfaces, coupling planes)
+/- 2kV, +/- 4kV, +/- 8kV,
and +/- 15kV (non-
conductive surfaces)