24.0 Care of Your Phototherapy Unit
24.1 Recommended Maintenance Schedule
Action
Frequency
Dusting of the unit and lamps
Once a month
Fully clean all internal reflectors,
lamps, and protective acrylic
Annually
Unit calibration
(Dosimetry Only)
Every 100 hours of use or once a year, whichever occurs
first and when lamps are replaced.
(Meters can be rented or purchased through Daavlin)
Power Output Measurement
(Timer Only)
Every 100 hours of use and when lamps are replaced.
(Meters can be rented or purchased through Daavlin)
Replace lamps
*UVB Approximately Every 300 hours of use.
*UVA Approximately Every 500 hours of use.
*Blue Approximately Every 500 hours of use
* Lamp life will vary significantly depending on average treatment time and other
environmental conditions.
24.2 Cleaning/Disinfection
24.2.1 General Cleaning
For general use, use a clean non-abrasive cloth, not paper towels, and a mild cleaning solution
such as Dawn Liquid Dishwashing Soap to gently wipe down the exterior of the device.
24.2.2 Low-Level Disinfection
To reduce micro-organisms on the patient-contacting surfaces, disinfect these surfaces between
uses of the device, including when the same patient has the device for use. For disinfection while
the same patient uses the device, we have tested several cleaners that do not degrade the Acrylic
and can be seen in Table 6 Daavlin Tested Cleaners.
Cleaner/Solution
Contact Time
Monk brand Wipes
Follow the contact time instructions
provided with the Monk brand Wipes
70% Isopropyl Alcohol
3 min
Table 6 Daavlin Tested Cleaners
1. Thoroughly wipe down the surfaces and allow contact time listed in Table 6 Daavlin Tested
Cleaners.
2. Allow to air dry and inspect for visible contaminants.
3. If contaminants remain repeat until no visible containments remain repeat steps 0 and 2.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Daavlin NeoLux 3 Series and is the answer not in the manual?

Daavlin NeoLux 3 Series Specifications

General IconGeneral
BrandDaavlin
ModelNeoLux 3 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals