PA Series Biofeedback and Stimulation System User Manual Basic Functions and Operations
24
counterclockwise to decrease the current intensity. Only during the electrical stimulation
output can the current be adjusted. The current cannot be adjusted during resting period.
4.3 Sensor/Probe/Electrodes Off
During the electrical stimulation, if any sensor/probe/electrode falls off or is not in good
contact with the skin, the electrical stimulation will stop automatically and “Channel has
fallen off” will be displayed on the interface. After replacement of the
sensor/probe/electrode, the system will identify whether it’s connected successfully, if so,
the treatment will go on.
WARNING
1 Please avoid contact between the electrodes and wounds or scars during the
treatment.
2 The adhesive electrodes should not be placed crossing over the heart.
CAUTION
1 Make sure the adhesive electrodes are in good contact with the skin.