PA Series Biofeedback and Stimulation System User Manual Maintenance
50
Chapter 6 Maintenance and Cleaning
6.1 Maintenance
6.1.1 Inspection
(1) Visual Inspection
Prior to using the system every time, perform the following inspections:
1 Check the system and accessories to see if there is any visible evidence of
damage that may affect the patient's or operator's safety.
2 Check all the outer cables, power socket and power cables for damage and
connection.
3 Check if the system functions properly.
If any damage is detected, stop using the system. Replace the damage part(s) or contact
the manufacturer for service before reusing it.
(2) Routine Inspection
The overall check of the system, including safety check and function check, should be
performed by qualified personnel every 12 months or according to local regulations, and
each time after service.
The equipment should undergo periodic safety testing to ensure proper patient isolation
from leakage currents. This should include leakage current measurement and insulation
testing. The recommended testing interval is once a year or as specified in the
institution's test and inspection protocol.
WARNING
Failure on the part of the responsible individual hospital or institution employing
the use of this equipment to implement a satisfactory maintenance schedule may
cause undue equipment failure and possible health hazards.
6.1.2 Daily Maintenance
Workstation:
Keep the workstation at least 5cm away from the wall for the sake of heat and air
dissipation.
To avoid the potential accidence resulted from unexpected power lost or unsteady
power supply, it is recommended to connect the system to a UPS (Uninterrupted
Power Supply).
This system is medical equipment. To protect the system from virus and crash,
third-party software without authorization should not be installed.
To ensure normal running of the operating system and the software, do not delete or
modify the files on the hard drive.