PA Series Biofeedback and Stimulation System User Manual Safety Guidance
4
4 SHOCK HAZARDDo not attempt to connect or disconnect a power cord with
wet hands. Make sure that your hands are clean and dry before touching a
power cord.
5 SHOCK HAZARD - The power receptacle must be a three-wire grounded outlet.
Never try to adapt the three-prong plug to fit a two-slot outlet. A hospital grade
outlet is required. If the outlet has only two slots, make sure that it is replaced
with a three-slot grounded outlet before attempting to operate the system.
6 Electromagnetic Interference Ensure that the environment in which the
device is installed is not subject to any source of strong electromagnetic
emissions, such as CT, radio transmitters, mobile phone base stations, etc.
7 Electromagnetic Interference Do not use mobile phones nearby the device,
otherwise it may interfere with device operation.
8 Do not touch the signal input or output connector and the patient
simultaneously.
9 Accessory equipment connected to the device must be certified according to
the respective IEC/EN standards (e.g. IEC/EN 60950 for data processing
equipment and IEC/EN 60601-1 for medical equipment). Furthermore all
configurations shall comply with the valid version of the system standard
IEC/EN 60601-1. If any question, please consult our technical service
department or your local distributor.
10 Connecting any accessory or equipment to the device makes a medical system.
In that case, additional safety measures should be taken during installation of
the system, and the system shall provide:
a) Within the patient environment, a level of safety comparable to that
provided by medical electrical equipment complying with IEC/EN 60601-1,
and
b) Outside the patient environment, the level of safety appropriate for
non-medical electrical equipment complying with other IEC or ISO safety
standards.
11 Do not exceed the maximum permitted load when using multiple portable
socket-outlets to supply the system. You can refer to the label or silk-screen
for the maximum permitted load.
12 SHOCK HAZARD Do not connect non-medical electrical equipment, which
has been supplied as a part of the system, directly to the wall outlet when the
non-medical equipment is intended to be supplied via a multiple portable
socket-outlet with a separating transformer.
WARNING
13 SHOCK HAZARD Do not connect electrical equipment, which has not been
supplied as a part of the system, to the multiple portable socket-outlet