HCEN12630 Revision 07.2015
Program version 1.xx
AK 98™ Dialysis Machine - Maintenance manual 185
by the local disinfection program. It requires integrated disinfection, together with the
water delivery system, or a manual disinfection procedure.
CAUTION!
Always use a peracetic disinfectant for both the machine and the inlet tube. Mixing
different types of disinfectants may be hazardous.
CAUTION!
Disinfectants may be toxic. Take note of necessary precautions before use. The
manufacturer’s instructions and recommendations shall be followed. Local
regulations regarding the utilization of the different chemicals shall be followed.
It is recommended to perform disinfection of the inlet part of the dialysis machine if the
machine is stored or switched off for a long period of time. The following procedure
can be used to disinfect the inlet part if integrated disinfection with the water treatment
system is not possible.
Procedure
1) Assemble the manual disinfection device as shown in Figure 8-1 “Manual
disinfection device” on page185. Contact your local Gambro Service
representative for information on how to obtain this device or the components to
assemble it.
Figure 8-1. Manual disinfection device
1. Adaptor for the connection to the
water inlet tube.
2. PVC tube 4.6/6.8 mm.
3. Non-return valve.
4. Disinfectant suction.
5. Silicone ‘T’ connector.
6. Nipple 3/6 mm.
7. PVC tube 3.5/5.5 mm.
8. 50 ml syringe.
2) Prepare a diluted peracetic (Dialox) solution of 2 ml peracetic/50 ml treated water
in a chemical resistant container (see Table 8-2 “Peracetic solution table” on