- 76 - eBike, eBike L, eBike EL 2018112-003 Rev J
Appendix A: Interfaces
command from PC feedback Description
(EKG unit) from eBike
Wxxx <CR> — sets nominal load value to xxx Watt
S <CR> — initiates BP measurement at rest, then starts the exercise
protocol with the initial load
s <CR> — initiates exercise protocol with the initial load, no BP
measurement at rest
B <CR> B xxx <CR> requests current nominal load
returns current load in W
D <CR> n xxx <CR> requests current RPM value
returns current RPM value
requests current systolic BP value
O <CR> O xxx <CR> returns current systolic BP value in mmHg
requests current diastolic BP value
U <CR> U xxx <CR> returns current diastolic BP value in mmHg
F <CR> - - - terminates exercise protocol
I <CR> erxxxP10Vyyy identifies the ergometer
<CR> The string contains the ergometer model „xxx“ and the software version.
Note: For compatibility with er900, the transferred string reads
„er900P10VV243“.
Transmission Protocol RS232 (PORT 1 and PORT 3): Digital EKG Unit